• <tr id='kHRIfu'><strong id='kHRIfu'></strong><small id='kHRIfu'></small><button id='kHRIfu'></button><li id='kHRIfu'><noscript id='kHRIfu'><big id='kHRIfu'></big><dt id='kHRIfu'></dt></noscript></li></tr><ol id='kHRIfu'><option id='kHRIfu'><table id='kHRIfu'><blockquote id='kHRIfu'><tbody id='kHRIfu'></tbody></blockquote></table></option></ol><u id='kHRIfu'></u><kbd id='kHRIfu'><kbd id='kHRIfu'></kbd></kbd>

    <code id='kHRIfu'><strong id='kHRIfu'></strong></code>

    <fieldset id='kHRIfu'></fieldset>
          <span id='kHRIfu'></span>

              <ins id='kHRIfu'></ins>
              <acronym id='kHRIfu'><em id='kHRIfu'></em><td id='kHRIfu'><div id='kHRIfu'></div></td></acronym><address id='kHRIfu'><big id='kHRIfu'><big id='kHRIfu'></big><legend id='kHRIfu'></legend></big></address>

              <i id='kHRIfu'><div id='kHRIfu'><ins id='kHRIfu'></ins></div></i>
              <i id='kHRIfu'></i>
            1. <dl id='kHRIfu'></dl>
              1. <blockquote id='kHRIfu'><q id='kHRIfu'><noscript id='kHRIfu'></noscript><dt id='kHRIfu'></dt></q></blockquote><noframes id='kHRIfu'><i id='kHRIfu'></i>

                电导率仪常见故障维修简析

                by starview

                很多人都知晓,电导率仪是测量水溶液电导率的数字显示式仪器,一个常见的问题,电导率仪在南方的梅雨季节使用时经常会出现显示数值不稳定,跳字的情况,此种故障如何解决呢?
                大家知道,电导率仪是数字显示仪器,此类仪器输入级电路由于采集的信号很微小,所以工作的稳定性对它很关键,在霉雨天气出现的电导率仪跳字故障可能和该级电路有关,如果仪器没有提供电路图,从实物了解到仪器的输入级由两只运放IC(LM324N、CA3140E)和外围元件组成,两只IC和印刷电路板都是由IC插座连接,此种插座连接方式虽然安装方便但存在着很大的弊端,就是它的接插端子所形成的接触电阻及热电势会随着环境湿度、温度、尘埃及接插压力等的变化而变化,而且此种变化是随机性质的,通常情况下,不可能完全予以补偿,此种情况一定会影响系统的稳定性,引起附加输出的“温漂"及“时漂"。
                另外,在信号的传输过程中经IC插座的活动弹性接点,都有可能会产生“隐患"。综合以上状况后,要先去除IC插座,再将两只运放 IC 直接焊在印刷电路板上。为避免外界有杂散的干扰信号从仪器的输入端进入而破坏输入级的稳定工作,再将输入端(测量所用的铂黑电极)和仪器输入级的连接线改为带屏蔽层与印刷电路板的“公共地"相连。
                采取了以上方法后,电导率仪器的工作就能稳定可靠了,能够完全解决仪器数值跳字的故障。